2010 Rev. 1
Section
1
38
D
URHAM
G
EO
S
LOPE
I
NDICATOR
SOILS
Soil Load Frames
S-611 Auto Load Frame
For CBR,
LBR,
UCC and
Triaxial
Test ing
ASTM
D 1883, D 2166
AASHTO
T 193, T 208
The S-611 Auto Load Frame is based on our time proven S-610
Load Frame (see p.30). When con nect ed to a PC, the S-611 Auto
Load Frame offers the ability to perform LBR, CBR and UCC tests
with a min i mum of operator in volve ment. Once the test is started,
the op er a tor may leave the test area. The com put er controls the
S-611, records load and defl ection readings, shows the test in
progress on an on-screen graph, au to mat i cal ly stops the test and
returns the platen to the start position for the next test. Separate
reporting software calculates results and provides fi nal reports
that can be printed.
The front-panel mounted selector switch allows the user to choose
the mode of operation between “manual” and “auto”. In manual
mode, the S-611 load frame operates the same as the S-610.
Compared to the S-610, the S-611 offers three major additions:
■ The platen will automatically retract to the lower home position
when a preset load and/or displacement condition is satisfi ed.
The platen will retract when a pre-set maximum displacement,
load, or load drop-off (indicating the test is completed) has
been reached or if sample failure has been detected.
■ New Smart Digital Indicators permit continuous transmission
of data. Programmable features include setting end of test
conditions and transmitting information to a com put er using
Windows
©
-based serial acquisition software called WinSAS™.
■ CBR and LBR software retrieves the data fi
le to produce data
report sheet, complete with load-deformation plots.
* See Standards Buyer's Guide at www.DGSI.info/3000.
S-611, shown
with optional
accessories
Software options on page 42.
Accessory Package for CBR Testing (S-61140)
Model Qty. Description
E-216 1 S-Type load Cell, 10,000 lbf (44.5 kN)
Or E-217 1 Or Cylindrical Type Load Cell, 10,000 lbf (44.5 kN)
E-311 1 Linear Displacement Transducer, 1 x 0.001 in
E-870 1 Dial Indicator Support Assembly
E-875 1 LDT Contact Bar
E-40508 1 RS-485 to RS-232 Converter
E-40520 1 Dual Smart Digital Indicators (5-Digit)
E-40521 1 WinSAS Collecting Software
E-41525 1 CBR Reporting Software
E-45010 1 Mounting Bracket for E-40520
S-61056 1 2-in Quick Adjust CBR / LBR Piston
600336 1 Centralizing Mold Fixture, S-611
600787 1 LDT Bracket, S-611
Specifi cations for S-611 Load Frame
Load Capacity 10,000 lbf (44.5 kN)
Speed Range for
Platen Movement
From 0.005 to 0.2 in/min. (0.127 to 5.08 mm/min)
in 0.001 in/min (0.0254 mm/min) steps.
Platen 6.25 in (159 mm) diameter
Piston 2 in (50 mm) diameter
Piston Travel 3 in (76 mm) maximum
Speed Control Microprocessor based electronic controller
Column Distance 11.75 in (298 mm)
Max. Daylight 32 in (812mm) without load measuring device
Strain Rods 1.5 in (38 mm) dia. Acme thread for adjustment
Cross Beam Adjustable, w/ center hole 0.75 in (19 mm)
Dimensions (WxDxH) 9 x 16 x 53 in (229 x 406 x 132 mm)
Ship. Wt. (w/o acc.) 200 lb (91 kg)
Ship. Vol. 4.75 ft³ (0.13 m³)
Accessory Package for UCC Testing (S-61130)
Model Qty. Description
E-875 1 LDT Contact Bar
E-212 1 S-Type Load Cell, 2,000 lbf capacity
E-281 1 Adapter for Load Cell to Cross Bar
E-311 1 Linear Displacement Transducer, 1 x 0.001 in
E-344 1 LDT Mounting Adapter
S-580 1 Upper Platen, 3 in diameter
S-581 1 Lower Platen, 3 in diameter
E-870 1 Dial Indicator Support Assembly
E-40508 1 RS-485 to RS-232 Converter
E-40520 1 Dual Smart Digital Indicators (5-Digit)
E-40521 1 WinSAS Collecting Software
E-41524 1 UCC Reporting Software
E-45010 1 Mounting Bracket for E-40520
S-611 Auto Load Frame, 10,000 lbf, 110 V, 60 Hz 160 lb
S-61110 Auto Load Frame, 10,000 lbf, 220 V, 60 Hz 160 lb
S-61115 Auto Load Frame, 10,000 lbf, 220-240 V, 50 Hz 180 lb
Accessories
S-61120 Accessory Package for LBR Testing (details on next page)
S-61130 Accessory Package for UCC Testing (details in next column)
S-61140 Accessory Package for CBR Testing (details in next column)
E-217 Cylindrical-Type Load Cell, 10,000 lbf (45 kN) (LBR Testing)